Data Output
Starrett Altissimo Operation Manual Page 15 CAT 2000-24
The Altissimo is equipped with an Electronic Data Output Port. Plug in the optional Data Output Cable to an
external Printer, Data Collector or Computer.
Communication Parameters:
Baud Rate: 4800
Data Bits: 8 (ASCII)
Stop Bits: 1
Parity: None
No. of Characters: 18 (including <CR> <LF>)
(1 = Negative Voltage, 0 = 0 voltage)
Data Format:
Inch Mode: <X> <X> <SP> < - > <X> <X> <X> < . > <X> <X> <X> <X> <X> <SP> <i> <n> <CR> <LF>
Metric Mode: <X> <X> <SP> < - > <X> <X> <X> <X> < . > <X> <X> <X> <X> <SP> <m> <m> <CR> <LF>
Prefix Sign Display Value Units
(SP if +)
Prefix Description
<SP> <SP> Normal Mode
DC Distance to Center
DF Distance to Last Feature
DH Height
DI Diameter
DP Distance Between Points
FS Force Setting
MN Minimum Reading
MX Maximum Reading
PB Preset Block Size
PR Preset
TR TIR Reading (Max Min)
Connector:
Ansley P/N 609-0653
Pin 1 GND
Pin 2 TX-DATA
Pin 3 RX-RQST
Pin 4 TX-CLK
Pin 5 Spare
Pin 6 Spare